Reconfigurable Microfluidic Magnetic Valve Arrays: Towards a Radiotherapy-Compatible Spheroid Culture Platform for the Combinatorial Screening of Cancer TherapiesCee® 200CBX
A Cee® 300X Precision Spin Coater was used to fabricate thin silicone membranes for a biomimetic microfluidic blood oxygenator. The membranes separate the device's vascular and oxygen channels, connecting legacy Cee® coating equipment with experimental gas-exchange device fabrication.
NuSil silicone was spin coated on a polished 300 mm silicon wafer at 1100 rpm for 60 s, with a 500 rpm/s ramp rate. The target membrane thickness was 50 µm. A trichlorosilane treatment enabled release from the wafer. The membranes rested at room temperature for 10 min before oven curing at 65 °C for at least 1.5 h. These are reported research conditions for this device. Source: Section 2.3, Membrane Fabrication.
The process details describe the cited work. Reported recipes and results depend on the authors’ materials, substrates, and equipment configuration. Materials recorded for this reference include NuSil Silicone.
Micromachines (MDPI). 2021. Toward Development of a Higher Flow Rate Hemocompatible Biomimetic Microfluidic Blood Oxygenator.
Open original source ↗Application summary prepared by Cee. The original source remains the authority for its methods and findings.